Vacu-Aide® Quiet Suction Unit

With more than 50% reduction in sound, the Vacu-Aide QSU is the quietest high flow/high suction portable unit on the market, making it a smart choice for patients from neonate to adult.

[READ MORE]

Vacu-Aide® QSU Suction Unit
For more than half a century, DeVilbiss Healthcare has provided unmatched safety and performance in the airway suction device category. With the Vacu-Aide QSU, DeVilbiss brings you the same reliable, portable operation in a significantly quieter unit.

With more than 50% reduction in sound, the Vacu-Aide QSU is the quietest high flow/high suction portable unit on the market, making it a smart choice for patients from neonate to adult.

Cleaning Instructions

Product Highlights

Vacuum adjustment allows for 50–550 mm Hg and a free flow of 27 liters/minute
Meets American Association for Respiratory Care (AARC) guidelines for neonatal, infant, child and adult homecare suctioning
Sealed lead acid battery for use up to one-hour operation
Can return a patient to a state of comfort quickly – Meets ISO 10079-1:2009 high vacuum, high flow classification
New, innovative container assembly features integrated bacteria filter
— Container and filter design maintains appropriate flow rate to meet ISO guidelines
Comes complete with internal battery, carrying case, 6′ tubing, AC/DC adapters and power cord
New Integrated gauge and flow regulator – Easy-to-turn knob conveniently placed around gauge for easy use and control
Manufactured and tested to ISO10079-1:1999 Standard for Suction Equipment, achieving outstanding performance in the following tests:
Electromagnetic Compatibility Test – Vacu-Aide does not interfere with or is not susceptible to interference from other equipment such as ventilators
Drop Test and Shock Vibration Test – Vacu-Aide is far less likely to be damaged when dropped or shipped
Excessive Temperature Test – Vacu-Aide effectively works in a wide range of environmental temperatures

AARC Guidelines for appropriate subatmospheric pressures:
Neonates: 60 – 80 mm Hg
Infants: 80 – 100 mm Hg
Children: 100 – 120 mm Hg
Adults: 100 – 150 mm Hg

Negative pressures should not exceed 150 mmHg as higher pressures have been shown to cause trauma, hypoxemia and atelectasis.

Source: American Association for Respiratory Care. Nasotracheal suctioning 2004 revision & update. Respir Care 2004 Sep;49(9): 1080-4 (62 references)